Previous : SYD-261A Automatic Pensky-Martens Closed-Cup Flash Point Tester
Next : SYD-0175 Distillate Fuel Oils Oxidation Stability Tester
- Product Introduction
Summary
This instrument is designed and manufactured as per standards GB/T 261-2008 Determination of Flash Point - Pensky-Martens closed cup method and ASTM D93 Standard Test Methods for Flash Point by Pensky-Martens Closed Cup Tester. It is used to make determination of the closed cup flash point of the petroleum products.
Main technical specifications
1. Power supply:AC( 220±10% )V, 50Hz.
2. Heating device:(1) The furnace is silicon carbide material. Power is 600W.
(2) The heating power is adjustable from 0 W to 600W.
3. Heating rate:Procedure A: (5~6)℃/min, Procedure B:(1~1.6)℃/min.
Automatically controlled
4. Stirring device:(1)Stirring motor: BYGH101 stepping motor
(2) Driving mode: flexible shaft
(3) Shaft size: 8mm×40mm
5. Stirring rate:Procedure A: (90~120)RPM, Procedure B: (250±10)RPM
6. Igniting device: (1) Igniting source: gas(or other civilian fuels,the same below)
(2) Electric ignition. Gas flame diameter:3.2mm~4.8mm
7. Temperature testing probe:RTD, PT100
8. Ambient temperature:≤ 35℃
9. Relative humidity:≤ 85%
10.Maximum power consumption:650W